1、IFC 010 Economical electromagnetic flow converterIFC 300 Full-feature electromagnetic flow converterKROHNE 05/20047.02499.21.00Technical Data SheetElectromagnetic flowmeters Variable area flowmeters Mass flowmeters Ultrasonic flowmeters Vortex flowmeters Flow controllers Level measuring instruments
2、Pressure and temperature Heat metering Communications technology Switches, counters, displays and recorders Engineering systems improved HMI extensive diagnostic functionalaty as default flexible with customized solutions for special requestsProduct features digital signal processing high-speed sign
3、al conversion system modular signal converterbasic version with extensive functionalityoptional modules to customise it to needs available for AC and DC voltage operation storage of calibration data in converter housingIFC 010The economical solutionThe electromagnetic flow converter IFC 010 is an ec
4、onomical solution for a wide range of applications.The IFC 010 can be used in combination with almost any OPTIFLUX.Typical use is found in those applications, where people require simple readout and data processing.Highlights basic and fully functional flow converter quick and easy to install and op
5、erate favourable price performance ratioProduct features digital signal processing high-speed signal conversion system modular signal converterbasic version without displayoptional: local display and control panel available for AC and DC voltage operation instrument diagnostics with MAGCHECKFields o
6、f application Industrieschemicalpaper galvanically is- Flow direction (forward or reverse) - Bidirectional measurement - Direction identified via status, pulse or curren- Continuous measurement of actual volume flow rate - Flow direction (forward or reverse) - Bidirectional measurement0.99,9 seconds
7、 (set in increments of 0,1) 0,2.99,9 seconds (set in increments of 0,1)configurable for all operating data; galvanicallystop output, output zero, reset counter, reset error- Direction identified via status, pulse or current outputConfigurable as measuring range identification for automatic range cha
8、nge, indicator of flow direction, overflow, errors, trip point or empty pipe indicationAll operating data configurable; galvanically isolated; HART communication* of Q100%, adjustable in 1% incrementsIFC 300 / IFC 0109IFC 300 / IFC 010m520 Vd 2032 0 pulses es per se msrom currt currentrrent outpcflo
9、w ratesolatednt outputy isolatedFULL-SCALE FLOWRATESv m/s0,3312v ft/s11040 DN mmminimumnominalmaximumDN inchminimumnominalmaximum 2,50,0050,0530,2121/100,020,230,93 40,0140,1360,5431/80,060,602,39 60,0310,3051,2211/40,131,345,38 100,0850,8483,3933/80,373,7314,94 150,1911,9097,6341/20,848,4033,61 200
10、,3393,39313,5723/41,4914,9459,76 250,5305,30121,20512,3323,3493,36 320,8698,68634,7441 1/43,8238,24152,97 401,35713,57254,2871 1/25,9859,75239,02 502,12121,20684,82329,3493,37373,47 653,58435,838143,3522 1/215,78157,79631,16 805,42954,288217,152323,90239,02956,09 1008,48284,822339,288437,35373,46149
11、3,84 12513,254132,537530,148558,35583,542334,17 15019,085190,851763,404684,03840,293361,17 20033,930339,3001357,2008149,391493,895975,57 25053,013530,1302120,52010233,412334,099336,37 30076,341763,4103053,64012336,123361,1913444,77 350103,9081039,0804156,32014457,494574,9318299,73 400135,7171357,170
12、5428,68016597,545975,4423901,76 450171,7651717,6506870,60018756,267562,5930250,34 500212,0582120,5808482,32020933,669336,6337346,53 600305,3703053,70012214,800241344,5013445,0453780,15 700415,6204156,20016624,800281829,9218299,2073196,79 800542,8805428,80021715,200322390,2323902,2995609,15 900687,06
13、06870,60027482,400363025,0330250,34121001,37 1000848,2208482,20033928,800403734,6037346,00149384,01 12001221,45012214,50048858,000485377,8853778,83215115,30 14001433,52014335,20057340,800566311,6063115,99252463,94 16002171,46021714,60086858,400649560,6595606,51382426,03 18002748,27027482,700109930,8
14、007212100,27121002,69484010,75 20003393,00033930,000135720,0008014938,93149389,29597557,18 22004105,50041055,000164220,0008818075,97180759,73723038,90 24004885,80048858,000195432,0009621511,53215115,30860461,20 26005733,90057339,000229356,00010425245,60252456,021009824,08 28006650,10066501,000266004
15、,00011229279,51292795,091171180,37 30007634,10076341,000305364,00012033611,93336119,311344477,23Q100% in m3/hQ100% in US gallons/minTabelle bitte engerIFC 300 / IFC 01010IFC 300 / IFC 010Connection box106771060,7 IFC 0101601051401,6 IFC 3002601552304,2* Dimensions and weight of body only, excl. cabl
16、e glandsApprox. weight in kg*lwhDimensions in mm*Connection box4,173,034,171,5 IFC 0106,304,135,513,5 IFC 30010,246,109,069,3* Dimensions and weight of body only, excl. cable glandsApprox. weight in lbs*Dimensions in inches*lhwDimensions and weightsIFC 010 CIFC 300 CIFC 300 FIFC 300 WIFC 010 CIFC 300 / IFC 01011IFC 300 / IFC 010