1、FUNCTION SETTINGS 1.Enter function setting mode. Press ON/OFF to turn the power off, Press ZERO first, then press ON/OFF, while keeping ZERO pressed, the display will show A_ON or A_OFF. 2. Selection of auto backlight mode. When the balance enter function setting menu. Press UNIT key, the display wi

2、ll show L_ON or L_OFF, press ZERO, selection display L_ON, this means auto backlight function active, press ZERO again, the display will show L_OFF, this means auto backlight function inactive, i.e. shut off backlight function. 3.Selection of auto shut off mode. When the balance enter function setti

3、ng menu. Press UNIT key, the display will show A_ON or A_OFF, press ZERO, selection display A_ON, this means auto shut off function active. Press ZERO again, the display will show A_OFF, this means auto shut off function inactive. 4. Return to weighing mode. Press ON/OFF to turn the power off, press

4、 ON/OFF again to turn on the power and the scale return to the weighing mode.COMPACT SCALEMRB SERIESINSTRUCTION MANUALSAFETY PRECAUTIONSAll safety messages are identified by the following the words “WARNING“ and “CAUTION“. These words mean the following: WARNINGImportant information to alert you to

5、a situation that might cause serious injury and damage to your property if instructions are not followed.CAUTIONImportant information that tells how to prevent damage to equipment.When using the scale, the following safety precautions should always be followed. WARNING Use only the AC adaptor with t

6、he scale. Other adaptor may cause damage. (AC adaptor is optional) CAUTIONAvoid installing the scale in direct sunlight, this may cause discoloration or malfunctions.Do not mix the type of batteries. Replace all other batteries at the same time.If the scale is not to be used for a long period of tim

7、e, remove all batteries from the battery compartment to avoid leakage which may cause damage to the instrument.Avoid overloading that could cause damage to the scale.Do not drop water in the scale that is not water-resistant. It causes damage, if inside of the scale is wet.Matter charged static elec

8、tricity could influence the weighing. Discharge the static electricity. As example, the method is to use the electrification prevention spray, to spray it to both side of weighing platform.1SPECIFICATIONSModel numberCapacityDivisionMRB -500 500g0.1gMRB -1200 1200g0.1gMRB- 6000 6000g1gMRB-10000 10000

9、g1gPlatter175160mmNet/gross weight700/880g Standard carton: 29.821.68.8(cm3)Package12 Units in one box: 6148.538 (cm3)Operating Temperature0-40(32-104)6AA Size Batteries or AC/DCPower sourceAdapter 9V/100mA6CALIBRATIONWhen to calibrateCalibration may be required when it is initially installed, if th

10、e scale is moved to a substantial distance. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations may occur.How to calibrate 1.Before entering calibration mode, press ON/OFF to turn the power on

11、for 1minutes. 2.Enter calibration mode Press ON/OFF to turn the power off. Press UNIT first, then Press ON/OFF, while keeping pressing UNIT, the display will show “CALu=”,press UNIT key to select calibration weight units(g or lb); pressZERO key to select calibration weight, the display will show “C

12、xxxxx”, X is flash digit. Press UNIT key the flash digit will move to right; Press TARE key to increase the flash digit.(X is the calibration weight which can be set according to your requirement). Press ZERO key to confirm the calibration weight, the display show “88888” and then AD value. 3.Calibr

13、ation by the weight Wait for the stable indicator to be displayed, Press UNIT the display will show “00SAVE“, after 2 seconds, the display will show first line calibration weight. 4.Place the calibration weight on platform. After the stable indicator is displayed, then press UNIT, the display will s

14、how “00SAVE“ and then return to A/D value, now calibration in finished. 5.Press ON/OFF, turn off the power, press ON/OFF turn on the power again, place full weight on the platform to make sure whether weighing is correct, if not, repeat steps 2-5. 6.Return to weighing mode Press ON/OFF to turn the p

15、ower off. Press ON/OFF again to turn on the power and the scale/balance return to the weighing mode.5DESCRIPTIONLEFT SIDEAC adaptorSocketTOPWeighing platform The arch face of platform is front. Correct the direction of the pan.DISPLAYTARE keyON/OFF key Turns the scalePower on/off.UNIT key Changes we

16、ighing unitsZERO key Sets display to zero or Subtracts weight of a container.BOTTOMBattery compartmentBattery compartment cover Batteries Batteries are not includedDISPLAYZERO /TAREindicator . Weighing unit indicatorSTABLELow battery indicator2PREPARATION1.Installing batteries/Connecting the AC adaptor BatteriesBattery