Study on Image Quality of Computed Radiography

上传人:li45****605 文档编号:29864617 上传时间:2018-01-25 格式:PDF 页数:6 大小:344.42KB
返回 下载 相关 举报
Study on Image Quality of Computed Radiography_第1页
第1页 / 共6页
Study on Image Quality of Computed Radiography_第2页
第2页 / 共6页
Study on Image Quality of Computed Radiography_第3页
第3页 / 共6页
Study on Image Quality of Computed Radiography_第4页
第4页 / 共6页
Study on Image Quality of Computed Radiography_第5页
第5页 / 共6页
点击查看更多>>
资源描述

《Study on Image Quality of Computed Radiography》由会员分享,可在线阅读,更多相关《Study on Image Quality of Computed Radiography(6页珍藏版)》请在金锄头文库上搜索。

1、Apr20 1 0,Volume 4,No4(Serial No29) Journal of Materials Science and Engineering,ISSN 1 9348959,USA N Study on Image Quality of Computed Radiography Gede Bayu Suparta,Nugroho Waskito and Lestari Sri Department ofPhysics,Faculty ofMathematics and Natural Sciences,Gadjah Mada University,tbg),akarta 55

2、281,Indonesia Received:January 13,2010Accepted:January 222010Published:April 25,2010 Abstract:This study examined the image quailty on Computed Radiography(CR)in conjunction with the proper radiation dose as indicated based on the set-up on the anodecathode voltage(V)and filament current(i)During th

3、is studya set of radiographic images of step wedge for various combinations of V and i was developedThe results showed that good image quality may be achieved by considering the effective V and i set upIncreasing V had a better effect to the image quality rather than increasing iThe Xray power setuD

4、 of 2540 kW with voltage 50 kV80 kV and current of 50 mA may be chosen depends on the targeted aluminum thickness equivalent to obtain good image quality at the acceptable low radiation dose Optimization setup based on aluminum(AI)thickness equivalent may lead to significant reduction on the radiati

5、on dose and power consumption Key words:Computed radiograph) radiation doseXrayradiology 1Introduction Computed Radiography(CR)is a new trend for radiography technique1】CR is very impressive due tO its ability in providing high resolution digital images, quantitative measurement,quick image producti

6、on and consistent performance on image developmentIt is also able to be processed and manipulated for better clarity2】CR has been applied on medical imaging, specifically on diagnostic radiography3The use of phosphor storage for developing temporary images has great benefit since the phosphor storag

7、e can be used many timesAlthough CR sounds promising,radiation safety still becomes a great concern4-8 In Indonesia,although CR has been accepted,the standard operation procedure for CR is still based on film radiography【9,1 0,especially in conjunction with the set-up on the anode-cathode voltage( a

8、nd the filament current i of the X-ray tubeV(in kV)and i(in mA)set up directly contributes to the X-ray radiation intensity that impinge the obj ect,so that the absorbed Xray energy in the objectHence,that Xray radiation Corresponding author:Gede Bayu Suparla(1 965-),male, associate professor,resear

9、ch fields:physicsphysics of imagingEmail:gbsupartaugmacid intensity( correlates to the absorbed energy dose(rad) or the equivalent dose(rem)in the object1 1141The relation between intensity l to the V and i setup can be described by Equation(1)【1 5 oc (1) Equation(1)shows that intensity I increases

10、by power of(20士01)when Vincreases,while it changes linearly when i increasesSince increasing V leads to increasing the mean energy,the penetration capability of the Xray also increasesHence the radiation dose also increasesConsidering the radiation safety,the combination between V and f has to be de

11、termined in accordance to the image quality being resultedThe V mainly effects to the power of penetration and so it is responsible to the image contrastThe i mainly effects to the image brightnessThe image sharpness mainly defined by the object position relative to the X-ray source and film screen(

12、CR casseue) This paper reported our study in a hospital in Indonesia related with the examination of the CR image quality based on the V and i setupThe aim of this research was to explore the possibility on reducing radiation dose to the patient through the V and i set-up in the CR implementation 2E

13、xperiment Study on Image Quality of Computed Radiography Experiment was carried out at one hospital closed to the Gadjah Mada UniversityThe CR facility comprised of an X-ray facility for providing X-ray radiation tracer to the object,a phosphor storage casseRe(PSC)to store a temporary image ofthe ob

14、ject, and a CR panel for converting the image on the PSC to a digital image of the objectThe CR images were taken and developed by an authorized operator in that hospita1The radiography process has followed the standard procedure that is regularly conducted by the radiographer The test was carried o

15、ut for a step wedge made of aluminum as the object targetIt is about 1 40 mm long and 40 mm width and it has 1 1 stepsThere was about 3 mm thick difference between stepsOn the radiography session,the PSC were laid on the object table,while the stepwedge was laid on the PSCThe object was set about 1

16、00 cm to the focal spot of Xray tubeSome images were taken for some V and i set-upsThe same PSC was used,hence the position of the PSC on the table and the position ofthe object on the PSC were set at about the same positions for all V and I setupsThe way the CR panel reads the PSC such that the orientations and the position of the casseRe were about the same The anode-cathode voltage V varied by 50 kV,6

展开阅读全文
相关资源
正为您匹配相似的精品文档
相关搜索

最新文档


当前位置:首页 > 学术论文 > 期刊/会议论文

电脑版 |金锄头文库版权所有
经营许可证:蜀ICP备13022795号 | 川公网安备 51140202000112号