《ECMO OXYGENATORSUniversity of Minnesota Medial …:ECMO氧明尼苏达大学医学…》由会员分享,可在线阅读,更多相关《ECMO OXYGENATORSUniversity of Minnesota Medial …:ECMO氧明尼苏达大学医学…(17页珍藏版)》请在金锄头文库上搜索。
1、ECMO OXYGENATORSSilicone Rubber Sheet Membranes Medtronic/AvecorRange of 5 sizes 0.8 - 4.5 M2FDA Approved for long term useMicroporous Hollow Fiber DevicesLimited duration of use do to plasma leak Medtronic Minimax - Polypropylene hollow fibers Infant/PediatricPlasma resistant fibers (PRF)Carmeda Co
2、ated (Heparin Bioactive surface)Medtronic Affinity - Polypropylene hollow fibers Large Peds and AdultsJostra Quadrox DPolymethylpentene fibersNo Plasma leakMICRO POROUS POLYPROPYLENE HOLLOW FIBERSMICRO POROUS POLYPROPYLENE HOLLOW FIBERSQUADROXOxygenation SystemQuadrox De-airingde-airingblood inQuadr
3、ox Rear viewgas ingas outwater inwater outluer portrecirculationblood outsampling porttemperature probeBlood FlowDe-airing membranepre chamberblood inblood outLuer portBlood Flowchamber Achamber BGas Distributionchamber Achamber BDistribution of Water Flowchamber Achamber BMembrane Structure7000 nmM
4、icroporous membranepore size max. 0.2 mbloodpathgaspath0,12 nmPolypropylene Membrane StructureQuadrox Diffusion MembraneTechnical DataQuadroxQuadrox DBlood flow rate0.5 7 l/min0.5 7 l/minPriming volume250 ml250 mlOxygenation membrane1.8 mpolypropylene1.8 mpolymethylpenteneMax. O2-transfer425 ml/min4
5、25 ml/minMax. CO2-transfer320 ml/min320 ml/minHeat exchanger0.6 m polyurethane0.6 m polyurethaneBlood Flow DirectionBottom-to-BottomBottom-to-BottomMax. water pressure14 psi (1bar)14 psi (1bar)Gas Exchange in the Artificial LungDiffusion controlled fixed variablesSurface area, membrane material, blood path thickness (boundary layer)Gradient drivenFio2 for O2 TransferNet Gas flow for CO2 TransferMembrane Structure7000 nmMicroporous membranepore size max. 0.2 mbloodpathgaspath0,12 nm