《tokeepthebalanceoperatingproperly,thehousingandplatform》由会员分享,可在线阅读,更多相关《tokeepthebalanceoperatingproperly,thehousingandplatform(4页珍藏版)》请在金锄头文库上搜索。
1、All the segments above belong to ASC . Of these, the unit can be set by ES-HA series electronic balances.4.9.3 Wiring method Computer Balance 1 Empty 1 2 RXD 2 RXD 3 TXD 3 TXD 4 4 5 Ground 5 Ground 6 6 7 7 8 8 9 Empty 9 4.9.4 Demonstrating communication software (1)The super end under WIN9X/NT/2000/
2、XP(2)The debugging assistant on (3)Self-arranging communication software 4.9.5 Order input ( as the order schedule at the right side ) Order Schedule5. Care and maintenance To keep the balance operating properly, the housing and platform should be kept clean and free from foreign material. If necess
3、ary, a cloth dampened with a mild detergent may be used. Keep calibration masses in a safe dry place. Unplug the AC Adapter when not in use. For long term storage, remove the batteries. 6. Error Codes The following list describes the various errors which may appear on the display and the suggested r
4、emedy. Display description Err 0 With the platform on, this signal indicates the platform is too light. Make it heavier properly. If this error still occurs, the sensor or the circuit board may be destroyed. The balance must be sent to the manufacturer to be serviced. Err 1 The sample being weighed
5、exceeds the capacity of the balance. If the error occurs when the sample is within the balance capacity, maybe the balance is not correctly calibrated.41.Description ES-HA series electronic balances do weighing and measuring with high precision loadcell, microcomputer and big liquid crystal display
6、screen. It has the advantage of simple operation, speedy weighing, accuracy, stability, and so on. It is widely used in factories, mines, institutions, agricultures, traffic, building, water conservancy, medicine, food, and so on. Its more applicable in business for weighing precious products. 2. Co
7、nstruction and installation Body Wind Cover Weighing Platform Level Adjust Foot Display Window Push Button 3. Packing list(1) Electronic balance 1 pc. (2) AC adapter 1 pc. (3) Operation Instruction 1 copy (4) Balance (Its division is 0.01g or higher) with one weight and a wind cover. 4. Operation4.1
8、 Turn onPut the round plug of AC adapter into the square hole at the back of the balance, then 1Order ButtonFunctionASCIIOOff4FH MCount4DH CCal43H TTare54H FMode46H UUnit55H PPercent50Hput AC adapter into an alternating outlet with two holes. Or install 6 alkaline batteries AA SIZE in the battery ho
9、lder at the bottom of the balance (pay attention to the Plus(+) and minus(-) ends). With no load on the platform, press . All the segments of the display will be shown, then the software version is displayed. When “0” is displayed, weighing can be done. 4.2 Turn off To turn off the balance, press wh
10、ile the balance is at work. 4.3 Weighing method Press to set the display to zero. Put the object to be weighted on the platform. Wait until the unit “g” appears and the reading is stable, then the weight can be read.4.4 Balance calibration When precision weighing is done, calibrate the balance as fo
11、llows: First take away the object from the platform, then press to set the display to zero. Press , displaying “SCALE” or “LINE”. “SCALE” means calibrating the balance sensitivity; “LINE” means calibrating the balance linear. Press to make sure (If press , the two kinds of calibration can be interch
12、angeable.)4.4.1 Sensitivity calibration When calibrating the balance sensitivity, two different weight values can be used. According to the shown calibrating weight value, press , the two calibrating weight values can be interchangeable, press to make sure. After that, the balance will display “0“.
13、In a few seconds, the balance will display the calibrating weight value. At this time, put the corresponding calibrating weight on the center of the platform according to the shown weight value. In a few seconds, the display will stop, and meanwhile the calibrating weight value will be shown, indica
14、ting that calibration is over. Normal weighing can be done.4.4.2 Linear calibration Being in the state of “LINE”, press , the balance will display “0”. In a few seconds, the balance will display 1/2 weight value of full capacity; according to the shown weight value, put the corresponding calibrating weight on the center of the platform. In a few seconds, the balance will display the full capacity weight value, according to the shown weight value, put the corr